A fume hood is a ventilated enclosure that draws hazardous vapours, gases and dust away from the person working at it. In Australia you will also hear it called a fume cupboard or fume cabinet. All three names describe the same basic idea: air flows in through the front opening, past the user, and carries contaminants away from their breathing zone.
Where that air goes next is the big difference. A ducted fume hood sends it outside the building through ductwork. A ductless fume hood cleans it through filters and returns it to the room. This guide explains how each works, what Australian Standards apply, and how to choose the right one for your lab.
What Is a Fume Hood?
Every fume hood has the same core parts:
- An enclosure made of chemical-resistant materials, with a work surface and side walls.
- A sash, a sliding glass front that you raise or lower. A lower sash gives better protection and higher air speed.
- Baffles at the back that spread airflow evenly across the work area.
- A fan that pulls air into the hood and away from you.
- An airflow indicator or alarm that warns you when airflow drops.
The key performance measure is face velocity, the speed of air moving into the hood across the sash opening.
Fume hood vs biosafety cabinet vs laminar flow cabinet
These three are often confused, and using the wrong one can be dangerous.
|
Cabinet |
Protects |
Use it for |
|---|---|---|
|
Fume hood (fume cupboard) |
The user, from chemical vapours and gases |
Chemicals, solvents, acids |
|
Biological safety cabinet |
The user, the sample and the environment, using HEPA-filtered air |
Infectious or biological materials |
|
Laminar flow cabinet |
The sample only; it blows clean air towards the user |
Sterile work like plating media. Never use it for chemicals |
How a Ducted Fume Hood Works
A ducted fume hood connects to an exhaust system. A fan, usually on the roof, pulls air from the room into the hood and pushes it out through ductwork to a discharge stack high above the building, where it disperses safely.
Because the air leaves the building, the room needs a steady supply of make-up air to replace it. Without enough make-up air, the hood cannot maintain its face velocity.
Common types of ducted fume cupboards include bench type, walk-in, variable air volume (VAV) and bypass designs. VAV cupboards automatically adjust airflow as you move the sash, keeping face velocity steady while saving energy.
Advantages
- Handles a wide range of chemicals, including unknown mixtures.
- Suitable for larger quantities, heating, boiling and reactions that release a lot of vapour.
- Suitable for highly toxic and carcinogenic chemicals.
- No filters to saturate or replace.
Disadvantages
- Higher installation cost, including ductwork, roof fan and building works.
- Fixed in one place.
- High running cost, because heated or cooled room air is exhausted continuously.
How a Ductless Fume Hood Works
A ductless fume hood, also called a recirculating or filtered fume cabinet, is a self contained unit. A built in fan draws air through filters, then returns the cleaned air to the room.
- Activated carbon filters adsorb chemical vapours. Different carbon grades are made for different chemical families, such as organic solvents, acids or ammonia.
- HEPA filters trap fine particles such as powders.
Advantages
- No ductwork, so installation is quick and much cheaper.
- Portable; many can be moved between rooms.
- Lower energy use, because conditioned room air is not exhausted outside.
- Useful where ducting is impossible, such as heritage buildings or rented spaces.
Limitations
- Filters only capture what they are designed for. Every ductless hood should have a written list of the chemicals and maximum quantities it is approved to handle.
- Filters saturate. Once the carbon is full, vapours pass straight through, known as breakthrough. Saturation monitoring and regular filter changes are essential.
- Not for every chemical. Ductless hoods are generally unsuitable for highly toxic or carcinogenic materials, chemicals with low boiling points (around 65 °C or less), large volumes, heating or changing chemical use.
- Ongoing filter costs, including disposal of used filters as hazardous waste.
For example, chloroform boils at 61 °C and is a possible carcinogen, so chloroform extractions belong in a ducted fume cupboard. (Link: our chloroform guide.)
Ducted vs Ductless: Side by Side
|
Factor |
Ducted fume hood |
Ductless fume hood |
|---|---|---|
|
Where the air goes |
Outside the building |
Filtered and returned to the room |
|
Chemical range |
Very broad, including unknowns |
Limited to the chemicals the filter is rated for |
|
Quantities |
Small to large |
Small |
|
Highly toxic or carcinogenic chemicals |
Suitable |
Generally not suitable |
|
Installation |
Ductwork, roof fan, building works |
Plug in, minimal installation |
|
Upfront cost |
Higher |
Lower |
|
Running cost |
High energy use |
Lower energy, plus filter replacement |
|
Mobility |
Fixed |
Often portable |
|
Australian Standard |
AS/NZS 2243.8 |
AS 2243.9 |
Australian Standards for Fume Cupboards
- AS/NZS 2243.8:2014 covers ducted fume cupboards. It requires an average face velocity of at least 0.5 m/s across the opening, with no more than 20% variation, and regular performance testing. Face velocity and smoke tests are typically done once a year.
- AS 2243.9:2025 covers recirculating (ductless) fume cabinets, including filter selection, fume absorption, face velocity and filter disposal.
Whichever you choose, ask your supplier for evidence that the unit meets the relevant standard and is commissioned and tested on site.
How to Choose: Ducted or Ductless?
Work through these questions:
- Which chemicals will you use, and will that list change? Known, stable and small quantities suit ductless. Varied or unknown chemicals need ducted.
- Are any highly toxic, carcinogenic or very volatile? If yes, choose ducted.
- Will you heat, boil or digest samples? Ducted is safer for large vapour loads.
- Can your building take ductwork and a roof fan? If not, a ductless hood may be the only practical option for low hazard work.
- What is your total budget over 10 years? Compare installation and energy for ducted against filter replacement for ductless.
- Who will manage filters, alarms and annual testing?
Typical picks:
|
Lab |
Usual choice |
|---|---|
|
University research or synthesis lab |
Ducted |
|
Pathology and analytical labs using many solvents |
Ducted |
|
School prep room with small amounts of known chemicals |
Ductless, if the chemical list is approved by the filter maker |
|
Weighing powders, staining or small scale low toxicity work |
Ductless |
|
Temporary or rented lab without ducting |
Ductless, for low hazard work only |
How to Use a Fume Hood Safely
- Check the airflow indicator before you start.
- Keep the sash as low as practical and never above the marked maximum height.
- Work at least 10 cm inside the front edge.
- Don't store chemicals or clutter inside; it disrupts airflow.
- Don't block the rear baffles.
- Move slowly in front of the hood; fast movements and passing traffic pull fumes out.
- Wear your PPE; a fume hood is not a replacement for goggles and gloves.
- Report alarms straight away and keep test records up to date.
Fun Facts
- A face velocity of 0.5 m/s is only about 1.8 km/h, slower than a gentle stroll. Fume hoods rely on smooth, steady airflow, not strong suction.
- Ducted fume hoods are among the largest energy users in a lab building. US studies estimate one can use as much energy in a year as several homes.
- "Fume cupboard" is the usual name in Australia, New Zealand and the UK, while "fume hood" is the American term.
Frequently Asked Questions
What is a fume cupboard?
A fume cupboard (fume hood) is a ventilated enclosure that draws hazardous fumes, vapours and dust away from the user and either exhausts them outside or filters them.
What is the difference between a ducted and ductless fume hood?
A ducted fume hood exhausts contaminated air outside through ductwork. A ductless fume hood passes air through carbon or HEPA filters and returns it to the room.
Are ductless fume hoods safe?
Yes, for the specific chemicals and quantities their filters are rated for, with regular saturation monitoring and filter changes. They are not suitable for highly toxic, carcinogenic or very volatile chemicals.
What face velocity should a fume cupboard have in Australia?
AS/NZS 2243.8 requires an average face velocity of at least 0.5 m/s, with no more than 20% variation.
How often should a fume hood be tested?
Face velocity and smoke testing are typically done at least once a year in Australia.
Can I use a laminar flow cabinet instead of a fume hood?
No. A laminar flow cabinet blows air towards the user and offers no protection from chemicals.